Comparison of PD BLADE with fat saturation (FS), PD FS and T2 3D DESS with water excitation (WE) in detecting articular knee cartilage defects

Thumbnail
Συγγραφέας
Lavdas, E.; Topalzikis, T.; Mavroidis, P.; Kyriakis, I.; Roka, V.; Kostopoulos, S.; Glotsos, D.; Zilidis, C.; Stathakis, S.; Tsagkalis, A.; Papanikolaou, N.; Batsikas, G.; Arvanitis, D. L.; Vassiou, K.
Ημερομηνία
2013
DOI
10.1016/j.mri.2013.06.007
Λέξη-κλειδί
1.5 T MRI
Knee MRI
BLADE proton density with fat saturation sequences
Cartilage defects
T2 3D DESS
STEADY-STATE DESS
ARTHROSCOPIC CORRELATION
HYALINE CARTILAGE
MENISCAL
TEARS
ECHO SEQUENCE
DUAL-ECHO
OSTEOARTHRITIS
APPEARANCE
MRI
ARTIFACTS
Radiology, Nuclear Medicine & Medical Imaging
Εμφάνιση Μεταδεδομένων
Επιτομή
The purpose of this study, is to compare the sequences: 1) proton density (PD) BLADE (BLADE is a PROPELLER-equivalent implementation of the Siemens Medical System) with fat saturation (FS) coronal (COR), 2) PD FS COR, 3) multi-planar reconstruction (MPR) with 3 mm slice thickness and 4) multi-planar reconstruction (MPR) with 1.5 mm slice thickness, both from the T2 3D-double-echo steady state (DESS) with water excitation (WE) sagittal (SAG), regarding their abilities to identify changes in the femorotibial condyle cartilage in knee MRI examinations. Thirty three consecutive patients with osteoarthritis (18 females, 15 males; mean age 56 years, range 37-71 years), who had been routinely scanned for knee examination using the previously mentioned image acquisition techniques, participated in the study. A quantitative analysis was performed based on the relative contrast (ReCON) measurements, which were taken both on normal tissues as well as on pathologies. Additionally, a qualitative analysis was performed by two radiologists. Motion and pulsatile flow artifacts were evaluated. The PD BLADE FS COR sequence produced images of higher contrast between Menisci and Cartilage, Fluid and Cartilage, Pathologies and Cartilage as well as of the Conspicuousness Superficial Cartilage and it was found to be superior to the other sequences (p < 0.001). The sequences T2 3D DESS 1.5 mm and T2 3D DESS 3 mm were significantly superior to the PD BLADE FS COR and the PD FS COR sequences in the visualization of Bone and Cartilage and the Conspicuousness Deep Surface Cartilage. This pattern of results is also confirmed by the quantitative analysis. PD FS BLADE sequences are ideal for the depiction of the cartilage pathologies compared to the conventional PD FS and 12 3D DESS sequences. (C) 2013 Elsevier Inc. All rights reserved.
URI
http://hdl.handle.net/11615/30213
Collections
  • Δημοσιεύσεις σε περιοδικά, συνέδρια, κεφάλαια βιβλίων κλπ. [19735]
htmlmap 

 

Πλοήγηση

Όλο το DSpaceΚοινότητες & ΣυλλογέςΑνά ημερομηνία δημοσίευσηςΣυγγραφείςΤίτλοιΛέξεις κλειδιάΑυτή η συλλογήΑνά ημερομηνία δημοσίευσηςΣυγγραφείςΤίτλοιΛέξεις κλειδιά

Ο λογαριασμός μου

ΣύνδεσηΕγγραφή (MyDSpace)
Πληροφορίες-Επικοινωνία
ΑπόθεσηΣχετικά μεΒοήθειαΕπικοινωνήστε μαζί μας
Επιλογή ΓλώσσαςΌλο το DSpace
EnglishΕλληνικά
htmlmap