• English
    • Ελληνικά
    • Deutsch
    • français
    • italiano
    • español
  • français 
    • English
    • Ελληνικά
    • Deutsch
    • français
    • italiano
    • español
  • Ouvrir une session
Voir le document 
  •   Accueil de DSpace
  • Επιστημονικές Δημοσιεύσεις Μελών ΠΘ (ΕΔΠΘ)
  • Δημοσιεύσεις σε περιοδικά, συνέδρια, κεφάλαια βιβλίων κλπ.
  • Voir le document
  •   Accueil de DSpace
  • Επιστημονικές Δημοσιεύσεις Μελών ΠΘ (ΕΔΠΘ)
  • Δημοσιεύσεις σε περιοδικά, συνέδρια, κεφάλαια βιβλίων κλπ.
  • Voir le document
JavaScript is disabled for your browser. Some features of this site may not work without it.
Tout DSpace
  • Communautés & Collections
  • Par date de publication
  • Auteurs
  • Titres
  • Sujets

Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study

Thumbnail
Auteur
Jaillette, E.; Zerimech, F.; De Jonckheere, J.; Makris, D.; Balduyck, M.; Durocher, A.; Duhamel, A.; Nseir, S.
Date
2013
DOI
10.1186/1471-2253-13-50
Sujet
Intubation
Polyurethane
Tracheal cuff
Microaspiration
Tracheal
injury
Pneumonia
CRITICALLY-ILL PATIENTS
VENTILATOR-ASSOCIATED PNEUMONIA
ENDOTRACHEAL-TUBE
AIRWAY PRESSURE
FLUID LEAKAGE
MICROASPIRATION
PREVENTION
VOLUME
FLOW
CARE
Anesthesiology
Afficher la notice complète
Résumé
Background: The primary objective of this study was to determine the efficiency of a pneumatic device in controlling cuff pressure (Pcuff) in patients intubated with polyurethane-cuffed tracheal tubes. Secondary objectives were to determine the impact of continuous control of Pcuff, and cuff shape on microaspiration of gastric contents. Methods: Prospective randomized controlled study. All patients requiring intubation and mechanical ventilation = 48 h were eligible. The first 32 patients were intubated with tapered polyurethane-cuffed, and the 32 following patients were intubated with cylindrical polyurethane-cuffed tracheal tubes. Patients randomly received 24 h of continuous control of Pcuff using a pneumatic device (Nosten r), and 24 h of routine care of Pcuff using a manometer. Target Pcuff was 25 cmH2O. Pcuff was continuously recorded, and pepsin was quantitatively measured in all tracheal aspirates during these periods. Results: The pneumatic device was efficient in controlling Pcuff (med [IQ] 26 [24, 28] vs 22 [20, 28] cmH2O, during continuous control of Pcuff and routine care, respectively; p = 0.017). In addition, percentage of patients with underinflation (31% vs 68%) or overinflation (53% vs 100%) of tracheal cuff, and percentage of time spent with underinflation (0.9 [0, 17] vs 14% [4, 30]) or overinflation (0 [0, 2] vs 32% [9, 54]) were significantly (p < 0.001) reduced during continuous control of Pcuff compared with routine care. No significant difference was found in microaspiration of gastric content between continuous control of Pcuff compared with routine care, or between patients intubated with tapered compared with cylindrical polyurethane-cuffed tracheal tubes. Conclusion: The pneumatic device was efficient in controlling Pcuff in critically ill patients intubated with polyurethane-cuffed tracheal tubes.
URI
http://hdl.handle.net/11615/28638
Collections
  • Δημοσιεύσεις σε περιοδικά, συνέδρια, κεφάλαια βιβλίων κλπ. [19735]
htmlmap 

 

Parcourir

Tout DSpaceCommunautés & CollectionsPar date de publicationAuteursTitresSujetsCette collectionPar date de publicationAuteursTitresSujets

Mon compte

Ouvrir une sessionS'inscrire
Help Contact
DepositionAboutHelpContactez-nous
Choose LanguageTout DSpace
EnglishΕλληνικά
htmlmap