Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study

Thumbnail
Autor
Jaillette, E.; Zerimech, F.; De Jonckheere, J.; Makris, D.; Balduyck, M.; Durocher, A.; Duhamel, A.; Nseir, S.
Fecha
2013
DOI
10.1186/1471-2253-13-50
Materia
Intubation
Polyurethane
Tracheal cuff
Microaspiration
Tracheal
injury
Pneumonia
CRITICALLY-ILL PATIENTS
VENTILATOR-ASSOCIATED PNEUMONIA
ENDOTRACHEAL-TUBE
AIRWAY PRESSURE
FLUID LEAKAGE
MICROASPIRATION
PREVENTION
VOLUME
FLOW
CARE
Anesthesiology
Mostrar el registro completo del ítem
Resumen
Background: The primary objective of this study was to determine the efficiency of a pneumatic device in controlling cuff pressure (Pcuff) in patients intubated with polyurethane-cuffed tracheal tubes. Secondary objectives were to determine the impact of continuous control of Pcuff, and cuff shape on microaspiration of gastric contents. Methods: Prospective randomized controlled study. All patients requiring intubation and mechanical ventilation = 48 h were eligible. The first 32 patients were intubated with tapered polyurethane-cuffed, and the 32 following patients were intubated with cylindrical polyurethane-cuffed tracheal tubes. Patients randomly received 24 h of continuous control of Pcuff using a pneumatic device (Nosten r), and 24 h of routine care of Pcuff using a manometer. Target Pcuff was 25 cmH2O. Pcuff was continuously recorded, and pepsin was quantitatively measured in all tracheal aspirates during these periods. Results: The pneumatic device was efficient in controlling Pcuff (med [IQ] 26 [24, 28] vs 22 [20, 28] cmH2O, during continuous control of Pcuff and routine care, respectively; p = 0.017). In addition, percentage of patients with underinflation (31% vs 68%) or overinflation (53% vs 100%) of tracheal cuff, and percentage of time spent with underinflation (0.9 [0, 17] vs 14% [4, 30]) or overinflation (0 [0, 2] vs 32% [9, 54]) were significantly (p < 0.001) reduced during continuous control of Pcuff compared with routine care. No significant difference was found in microaspiration of gastric content between continuous control of Pcuff compared with routine care, or between patients intubated with tapered compared with cylindrical polyurethane-cuffed tracheal tubes. Conclusion: The pneumatic device was efficient in controlling Pcuff in critically ill patients intubated with polyurethane-cuffed tracheal tubes.
URI
http://hdl.handle.net/11615/28638
Colecciones
  • Δημοσιεύσεις σε περιοδικά, συνέδρια, κεφάλαια βιβλίων κλπ. [19735]
htmlmap 

 

Listar

Todo DSpaceComunidades & ColeccionesPor fecha de publicaciónAutoresTítulosMateriasEsta colecciónPor fecha de publicaciónAutoresTítulosMaterias

Mi cuenta

AccederRegistro
Help Contact
DepositionAboutHelpContacto
Choose LanguageTodo DSpace
EnglishΕλληνικά
htmlmap