| dc.description.abstract | The aim of the present study was to compare the aortic systolic blood pressure (aSBP), heart-rate-adjusted augmentation index (AIx75), and pulse wave velocity (PWV) obtained using the Mobil-O-Graph (IEM, Stolberg, Germany) and SphygmoCor (AtCor, Sydney, Australia) devices in patients receiving peritoneal dialysis (PD).After a 10-minute rest in the supine position, the Mobil-O-Graph and SphygmoCor devices were applied in randomized order in 27 consecutive PD patients. The agreement between the measurements produced by the Mobil-O-Graph and SphygmoCor devices was explored using Bland-Altman analysis.The Mobil-O-Graph-derived aSBP, AIx75, and PWV did not differ from the same measurements obtained with SphygmoCor (aSBP: 120.5 ± 18.2 mmHg vs. 124.4 ± 19.0 mmHg, p = 0.438; AIx75: 27.0% ± 12.4% vs. 24.5% ± 10.6%, p = 0.428; PWV: 9.5 ± 2.1 m/s vs. 10.1 ± 3.1 m/s, p = 0.397). The slight difference in the estimation of aSBP is possibly explained by the difference in brachial SBP used for the calibration of the devices (131.0 ± 20.6 mmHg vs. 134.5 ± 19.7 mmHg, p = 0.525). Mobil-O-Graph-derived measurements correlated strongly with paired measurements obtained with the SphygmoCor device. Bland-Altman plots showed no evidence of asymmetry and a wide range of agreement between the two devices.Our study shows acceptable agreement between Mobil-O-Graph and SphygmoCor in the estimation of arterial stiffness indices in PD patients. Accordingly, the Mobil-O-Graph device accurately performs aortic ambulatory blood pressure monitoring in this population. | en |
| dc.language.iso | en | en |
| dc.source | Advances in peritoneal dialysis. Conference on Peritoneal Dialysis | en |
| dc.source.uri | https://www.scopus.com/inward/record.uri?eid=2-s2.0-85057209371&partnerID=40&md5=25d0a8f02022ac1eaca5c8ba9cce5772 | |
| dc.subject | arterial stiffness | en |
| dc.subject | blood pressure | en |
| dc.subject | blood pressure monitoring | en |
| dc.subject | human | en |
| dc.subject | oscillometry | en |
| dc.subject | peritoneal dialysis | en |
| dc.subject | pulse wave | en |
| dc.subject | validation study | en |
| dc.subject | Blood Pressure | en |
| dc.subject | Blood Pressure Monitoring, Ambulatory | en |
| dc.subject | Humans | en |
| dc.subject | Oscillometry | en |
| dc.subject | Peritoneal Dialysis | en |
| dc.subject | Pulse Wave Analysis | en |
| dc.subject | Vascular Stiffness | en |
| dc.subject | NLM (Medline) | en |
| dc.title | Accuracy of a Newly-Introduced Oscillometric Device for the Estimation of Arterial Stiffness Indices in Patients on Peritoneal Dialysis: A Preliminary Validation Study | en |
| dc.type | journalArticle | en |