Prospective PET image quality gain calculation method by optimizing detector parameters

Thumbnail
Auteur
Theodorakis L., Loudos G., Prassopoulos V., Kappas C., Tsougos I., Georgoulias P.
Date
2015
Language
en
DOI
10.1097/MNM.0000000000000381
Sujet
Article
computer assisted emission tomography
computer simulation
cost effectiveness analysis
digital imaging and communications in medicine
energy
image quality
Monte Carlo method
PET-CT scanner
physics
quantitative analysis
radiation attenuation
radiological parameters
signal noise ratio
image processing
positron emission tomography
procedures
signal noise ratio
theoretical model
Image Processing, Computer-Assisted
Models, Theoretical
Monte Carlo Method
Positron-Emission Tomography
Signal-To-Noise Ratio
Lippincott Williams and Wilkins
Afficher la notice complète
Résumé
Background Lutetium-based scintillators with highperformance electronics introduced time-of-flight (TOF) reconstruction in the clinical setting. Let G′ be the total signal to noise ratio gain in a reconstructed image using the TOF kernel compared with conventional reconstruction modes. G′ is then the product of G1 gain arising from the reconstruction process itself and (n.1) other gain factors (G2, G3,...Gn) arising from the inherent properties of the detector. Methods We calculated G2 and G3 gains resulting from the optimization of the coincidence and energy window width for prompts and singles, respectively. Both quantitative and image-based validated Monte Carlo models of Lu2SiO5 (LSO) TOF-permitting and Bi4Ge3O12 (BGO) TOFnonpermitting detectors were used for the calculations. Results G2 and G3 values were 1.05 and 1.08 for the BGO detector and G3 was 1.07 for the LSO. A value of almost unity for G2 of the LSO detector indicated a nonsignificant optimization by altering the energy window setting. G′ was found to be ~1.4 times higher for the TOF-permitting detector after reconstruction and optimization of the coincidence and energy windows. Conclusion The method described could potentially predict image noise variations by altering detector acquisition parameters. It could also further contribute toward a long-lasting debate related to cost-efficiency issues of TOF scanners versus the non-TOF ones. Some vendors re-engage nowadays to non-TOF product line designs in an effort to reduce crystal costs. Therefore, exploring the limits of image quality gain by altering the parameters of these detectors remains a topical issue. © 2015 Wolters Kluwer Health, Inc. All rights reserved.
URI
http://hdl.handle.net/11615/79674
Collections
  • Δημοσιεύσεις σε περιοδικά, συνέδρια, κεφάλαια βιβλίων κλπ. [19735]
htmlmap 

 

Parcourir

Tout DSpaceCommunautés & CollectionsPar date de publicationAuteursTitresSujetsCette collectionPar date de publicationAuteursTitresSujets

Mon compte

Ouvrir une sessionS'inscrire
Help Contact
DepositionAboutHelpContactez-nous
Choose LanguageTout DSpace
EnglishΕλληνικά
htmlmap