dc.description.abstractPneumatic tube system (PTS) is a widely used method of transporting blood samples in hospitals. The aim of this study was to evaluate the effects of the PTS transport in certain routine laboratory parameters as it has been implicated with hemolysis. A systematic review and a meta-analysis were conducted. PubMed and Scopus databases were searched (up until November 2016) to identify prospective studies evaluating the impact of PTS transport in hematological, biochemical and coagulation measurements. The random-effects model was used in the meta-analysis utilizing the mean difference (MD). Heterogeneity was quantitatively assessed using the Cohran's Q and the I2 index. Subgroup analysis, meta-regression analysis, sensitivity analysis, cumulative meta-analysis and assessment of publication bias were performed for all outcomes. From a total of 282 studies identified by the searching procedure, 24 were finally included in the meta-analysis. The meta-analysis yielded statistically significant results for potassium (K) [MD=0.04 mmol/L; 95% confidence interval (CI)=0.015-0.065; p=0.002], lactate dehydrogenase (LDH) (MD=10.343 U/L; 95% CI=6.132-14.554; p<10-4) and aspartate aminotransferase (AST) (MD=1.023 IU/L; 95% CI=0.344-1.702; p=0.003). Subgroup analysis and random-effects meta-regression analysis according to the speed and distance of the samples traveled via the PTS revealed that there is relation between the rate and the distance of PTS with the measurements of K, LDH, white blood cells and red blood cells. This meta-analysis suggests that PTS may be associated with alterations in K, LDH and AST measurements. Although these findings may not have any significant clinical effect on laboratory results, it is wise that each hospital validates their PTS. © 2017 Walter de Gruyter GmbH, Berlin/Boston.en
dc.language.isoenen
dc.sourceClinical Chemistry and Laboratory Medicineen
dc.source.urihttps://www.scopus.com/inward/record.uri?eid=2-s2.0-85032974140&doi=10.1515%2fcclm-2017-0008&partnerID=40&md5=0aacd12489b21a9edec55abe84377866
dc.subjectaspartate aminotransferaseen
dc.subjecthemoglobinen
dc.subjectlactate dehydrogenaseen
dc.subjectpotassiumen
dc.subjectaspartate aminotransferaseen
dc.subjectlactate dehydrogenaseen
dc.subjectpotassiumen
dc.subjectbasophilen
dc.subjectblood biochemistryen
dc.subjectblood samplingen
dc.subjectcomparative studyen
dc.subjectcore laboratoryen
dc.subjecteosinophil counten
dc.subjecterythrocyte counten
dc.subjecterythrocyte sedimentation rateen
dc.subjecthematocriten
dc.subjecthematological parametersen
dc.subjecthemolysisen
dc.subjecthumanen
dc.subjectlymphocyte counten
dc.subjectmean corpuscular hemoglobinen
dc.subjectmean corpuscular volumeen
dc.subjectmeta analysisen
dc.subjectmonocyteen
dc.subjectpartial thromboplastin timeen
dc.subjectpriority journalen
dc.subjectprothrombin timeen
dc.subjectquantitative analysisen
dc.subjectred blood cell distribution widthen
dc.subjectreticulocyte counten
dc.subjectReviewen
dc.subjectsystematic reviewen
dc.subjectthrombocyte counten
dc.subjecttraffic and transporten
dc.subjectvelocityen
dc.subjectblooden
dc.subjectblood samplingen
dc.subjecthemolysisen
dc.subjecthospital laboratoryen
dc.subjectmetabolismen
dc.subjectbiochemistryen
dc.subjectblood clotting parametersen
dc.subjectblood examinationen
dc.subjecterythrocyteen
dc.subjectlaboratoryen
dc.subjectleukocyteen
dc.subjectmeasurementen
dc.subjectmedical proceduresen
dc.subjectMedlineen
dc.subjectoutcome assessmenten
dc.subjectprospective studyen
dc.subjectpublishingen
dc.subjectScopusen
dc.subjectsensitivity analysisen
dc.subjectAspartate Aminotransferasesen
dc.subjectBlood Specimen Collectionen
dc.subjectHemolysisen
dc.subjectHumansen
dc.subjectL-Lactate Dehydrogenaseen
dc.subjectLaboratories, Hospitalen
dc.subjectPotassiumen
dc.subjectWalter de Gruyter GmbHen
dc.titleThe impact of pneumatic tube system on routine laboratory parameters: A systematic review and meta-analysisen
dc.typeotheren


Αρχεία σε αυτό το τεκμήριο

ΑρχείαΜέγεθοςΤύποςΠροβολή

Δεν υπάρχουν αρχεία που να σχετίζονται με αυτό το τεκμήριο.

Αυτό το τεκμήριο εμφανίζεται στις ακόλουθες συλλογές

Εμφάνιση απλής εγγραφής