Show simple item record

dc.creatorTsougos, I. M.en
dc.creatorTheodorou, K.en
dc.creatorBazioglou, M. A.en
dc.creatorStathakis, S.en
dc.creatorKappas, C.en
dc.date.accessioned2015-11-23T10:52:24Z
dc.date.available2015-11-23T10:52:24Z
dc.date.issued2004
dc.identifier.issn11070625
dc.identifier.urihttp://hdl.handle.net/11615/34072
dc.description.abstractPurpose: To compare Monte Carlo simulation with conventional dosimetry techniques for stereotactic radiotherapy (SRT), since accurate dosimetry of narrow photon beams is very complicated and has often been questioned, mainly due to the lack of lateral electronic equilibrium and uncertainty in beam energy in terms of steep dose gradients. Materials and methods: In this work a Monte Carlo (MC, EGS4) simulation for dosimerty study was performed for the 6MV home made SRT unit of the University Hospital of Patras (Hellas). The results were compared with conventional small field dosimetry techniques such as ionization chamber, TLD's, and films (conventional and radiochromic). Hence, a comparison of many of the dosimetric techniques currently being used in small field dosimetry was attempted. Results: It was shown that all techniques are in reasonable agreement (within ± 2%) and that Monte Carlo can be used as a reliable reference for the dosimetry of the SRT beams, especially where lateral electronic equilibrium does not exist, as long as accurate simulation can be achieved. Conclusion: This study is only limited by the insurance of accurate simulation of the linear accelerator, which can be a difficult task since it is limited by the availability of the manufacturer's designs and the availability of computers and computer time for adequate runs, but it could become a useful tool for Monte Carlo simulations, as it contains detailed analysis of the run parameters and component modules selection. © 2004 Zerbinis Medical Publications.en
dc.sourceJournal of B.U.ON.en
dc.source.urihttp://www.scopus.com/inward/record.url?eid=2-s2.0-14044261250&partnerID=40&md5=d9a1cba1a0a38b850ce2ec5300a04ad8
dc.subjectMonte Carloen
dc.subjectRadiosurgeryen
dc.subjectSmall field dosimetryen
dc.subjectSterotactic radiotherapyen
dc.subjectaccuracyen
dc.subjectarticleen
dc.subjectbeam therapyen
dc.subjectcomparative studyen
dc.subjectcomputeren
dc.subjectcontrolled studyen
dc.subjectdosimetryen
dc.subjectelectronicsen
dc.subjectGreeceen
dc.subjectionization chamberen
dc.subjectlinear acceleratoren
dc.subjectMonte Carlo methoden
dc.subjectphotonen
dc.subjectradiation energyen
dc.subjectradiation fielden
dc.subjectradiotherapyen
dc.subjectreliabilityen
dc.subjectstandarden
dc.subjectstereotaxisen
dc.subjectuniversity hospitalen
dc.titleA comparison of Monte Carlo simulation with experimental dosimetric techniques for a 6 MV stereotactic radiotherapy uniten