dc.description.abstractBackground and purpose: Low megavoltage photon beams are often the treatment choice in radiotherapy when low density heterogeneities are involved, because higher energies show some undesirable dosimetric effects. This work is aimed at investigating the effects of different energy selection for low density tissues. Patients and methods: BEAMnrc was used to simulate simple treatment set-ups in a simple and a CT reconstructed lung phantom and an air-channel phantom. The dose distribution of 6, 15 and 20 MV photon beams was studied using single, AP/PA and three-field arrangements. Results: Our results showed no significant changes in the penumbra width in lung when a pair of opposed fields were used. The underdosage at the anterior/posterior tumor edge caused by the dose build-up at the lung-tumor interface reached 7% for a 5 X 5 cm AP/PA set-up. Shrinkage of the 90% isodose volume was noticed for the same set-up, which could be rectified by adding a lateral field. For the CT reconstructed phantom, the AP/PA set-up offered better tumor coverage when lower energies were used but for the three field set-up, higher energies resulted to better sparing of the lung tissue. For the air-channel set-up, adding an opposed field reduced the penumbra width. Using higher energies resulted in a 7% cold spot around the air-tissue interface for a 5 X 5 cm field. Conclusions: The choice of energy for treatment in the low density areas is not a straightforward decision but depends on a number of parameters such as the beam set-up and the dosimetric criteria. Updated calculation algorithms should be used in order to be confident for the choice of energy of treatment (c) 2006 Elsevier Ireland Ltd. All rights reserved.en
dc.sourceRadiotherapy and Oncologyen
dc.source.uri<Go to ISI>://WOS:000237911500020
dc.subjectradiotherapyen
dc.subjectheterogeneitiesen
dc.subjectenergy choiceen
dc.subject18 MV PHOTONSen
dc.subjectX-RAY-BEAMen
dc.subjectELECTRON-TRANSPORTen
dc.subjectDOSE CALCULATIONSen
dc.subjectAIR-CAVITYen
dc.subjectLUNGen
dc.subjectRADIOTHERAPYen
dc.subjectCONVOLUTIONen
dc.subjectALGORITHMSen
dc.subjectLESIONSen
dc.subjectOncologyen
dc.subjectRadiology, Nuclear Medicine & Medical Imagingen
dc.titleMonte Carlo dosimetric evaluation of high energy vs low energy photon beams in low density tissuesen