A finite element model of the Ilizarov fixator system

Thumbnail
Author
Toumanidou, Th; Spyrou, L. A.; Aravas, N.
Date
2011
DOI
10.1109/IWBE.2011.6079016
Keyword
3D finite element model
Axial stiffness
Biomechanical analysis
Configuration parameters
Finite element models
Frame stiffness
Ilizarov fixator
Mechanical characteristics
Non-linear stiffness
Parametric analysis
Ring diameter
Wire diameter
Biomechanics
Biomedical engineering
Stiffness
Three dimensional
Wire
Finite element method
Metadata display
Abstract
Mechanical characteristics of orthopaedic fixators, such as stiffness and stability, directly influence the mechanobiological environment in which the bone is healed. A comprehensive understanding of the sensitivity of the Ilizarov Apparatus (IA) to configuration parameters is the key to its biomechanical analysis. In this study, detailed 3D Finite Element (FE) models of two-ring and four-ring configurations of the IA are developed and studied with respect to axial stiffness. A parametric analysis is carried out on the four-ring configuration and the effects of the ring diameter, the wire diameter, and pretension level on the frame stiffness are studied. Both configurations are found to exhibit a non-linear stiffness behavior under axial compression. It is found that the addition of levels of fixations, i.e., rings and wires, augments fixator's stability. The analysis shows that decreasing the rings size and increasing the wires diameter and pretension levels provides greater frame stiffness. The results are in good agreement with the findings reported in literature. © 2011 IEEE.
URI
http://hdl.handle.net/11615/33684
Collections
  • Δημοσιεύσεις σε περιοδικά, συνέδρια, κεφάλαια βιβλίων κλπ. [19735]
htmlmap 

 

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

LoginRegister (MyDspace)
Help Contact
DepositionAboutHelpContact Us
Choose LanguageAll of DSpace
EnglishΕλληνικά
htmlmap