dc.description.abstractBackground: To determine the repeatability and reproducibility of optical coherence tomography (OCT) Spectralis retinal thickness measurements in diabetic patients with clinically significant macular edema (CSME). Methods: Twelve eyes of 12 volunteers (without macular pathology-control group) and 21 eyes of 21 diabetic patients with CSME were included in the study. Reproducibility, repeatability, intraclass correlation coefficients (ICCs) and intrasession correlation coefficients were tested with 20 × 15 degree raster scans consisting of 19 high-resolution line scans that were repeated three times by two experienced examiners. Results: In the control group, examining all regions, coefficient of repeatability was less than 1.1%, while coefficient of reproducibility was less than 2.2%. In diabetic patients, examining all regions, coefficient of repeatability was less than 2.6%, while coefficient of reproducibility was less than 2.4%. ICCs were, respectively, greater than or equal to 0.98 in the control group and 0.99 in diabetic patients. Intrasession coefficients of variation were less than 0.4% in the control group and less than 0.5% in diabetic patients. Conclusion: Retinal thickness measurements are repeatable and reproducible with OCT Spectralis in both the control group and diabetic patients. The results indicate that a change in central subfield thickness exceeding 12μm and 3% in the diabetic patients is likely to be real. © 2013 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.en
dc.sourceCurrent Eye Researchen
dc.source.urihttp://www.scopus.com/inward/record.url?eid=2-s2.0-84877603098&partnerID=40&md5=c8def21546f6e4850e4eca85e7c7269c
dc.subjectHeidelberg spectralisen
dc.subjectOptical coherence tomographyen
dc.subjectRepeatabilityen
dc.subjectReproducibilityen
dc.subjectarticleen
dc.subjectclinical articleen
dc.subjectcontrolled studyen
dc.subjectcorrelation coefficienten
dc.subjectdiabetes mellitusen
dc.subjectdiabetic macular edemaen
dc.subjectdiabetic patienten
dc.subjecthumanen
dc.subjectmeasurement repeatabilityen
dc.subjectpriority journalen
dc.subjectprospective studyen
dc.subjectretinal thicknessen
dc.subjectspectral domain optical coherence tomographyen
dc.subjectAdulten
dc.subjectDiabetic Retinopathyen
dc.subjectFemaleen
dc.subjectHumansen
dc.subjectMacular Edemaen
dc.subjectMaleen
dc.subjectProspective Studiesen
dc.subjectReproducibility of Resultsen
dc.subjectRetinaen
dc.subjectSoftwareen
dc.subjectTomography, Optical Coherenceen
dc.subjectVisual Acuityen
dc.titleRepeatability and reproducibility of retinal thickness measurements in diabetic patients with spectral domain optical coherence tomographyen