Repository of intra- and inter-run variations of quantitative autoantibody assays: A European multicenter study
dc.creator | Dragon-Durey M.-A., Bizzaro N., Senant M., Andreeva H., Bogdanos D.P., Bonroy C., Bossuyt X., Eriksson C., Fabien N., Heijnen I., Herold M., Musset L., Kuhi L., Lopez-Hoyos M., Berki T., Roozendaal C., Sack U., Sundic T., Taylor L., Kuna A.T., Damoiseaux J. | en |
dc.date.accessioned | 2023-01-31T07:59:19Z | |
dc.date.available | 2023-01-31T07:59:19Z | |
dc.date.issued | 2022 | |
dc.identifier | 10.1515/cclm-2022-0411 | |
dc.identifier.issn | 14346621 | |
dc.identifier.uri | http://hdl.handle.net/11615/73466 | |
dc.description.abstract | Objectives: No reference data are available on repositories to measure precision of autoantibody assays. The scope of this study was to document inter- and intra-run variations of quantitative autoantibody assays based on a real-world large international data set. Methods: Members of the European Autoimmunity Standardisation Initiative (EASI) group collected the data of intra- and inter-run variability obtained with assays quantifying 15 different autoantibodies in voluntary participating laboratories from their country. We analyzed the impact on the assay performances of the type of immunoassay, the number of measurements used to calculate the coefficient of variation (CVs), the nature and the autoantibody level of the internal quality control (IQC). Results: Data were obtained from 64 laboratories from 15 European countries between February and October 2021. We analyzed 686 and 1,331 values of intra- and inter-run CVs, respectively. Both CVs were significantly dependent on: the method of immunoassay, the level of IQC with higher imprecision observed when the antibody levels were lower than 2-fold the threshold for positivity, and the nature of the IQC with commercial IQCs having lower CVs than patients-derived IQCs. Our analyses also show that the type of autoantibody has low impact on the assay' performances and that 15 measurements are sufficient to establish reliable intra- and inter-run variations. Conclusions: This study provides for the first time an international repository yielding values of intra- and inter-run variation for quantitative autoantibody assays. These data could be useful for ISO 15189 accreditation requirements and will allow clinical diagnostic laboratories to assure quality of patient results. © 2022 Walter de Gruyter GmbH, Berlin/Boston. | en |
dc.language.iso | en | en |
dc.source | Clinical Chemistry and Laboratory Medicine | en |
dc.source.uri | https://www.scopus.com/inward/record.uri?eid=2-s2.0-85133542085&doi=10.1515%2fcclm-2022-0411&partnerID=40&md5=38e1e885d620b9c7245749f487423dcf | |
dc.subject | autoantibody | en |
dc.subject | beta2 glycoprotein 1 antibody | en |
dc.subject | cardiolipin antibody | en |
dc.subject | cyclic citrullinated peptide antibody | en |
dc.subject | double stranded DNA antibody | en |
dc.subject | gliadin antibody | en |
dc.subject | immunoglobulin G | en |
dc.subject | immunoglobulin M | en |
dc.subject | protein glutamine gamma glutamyltransferase antibody | en |
dc.subject | rheumatoid factor | en |
dc.subject | thyroglobulin antibody | en |
dc.subject | thyrotropin receptor antibody | en |
dc.subject | autoantibody | en |
dc.subject | accreditation | en |
dc.subject | addressable laser bead immunoassay | en |
dc.subject | Article | en |
dc.subject | Austria | en |
dc.subject | Belgium | en |
dc.subject | benchmarking | en |
dc.subject | controlled study | en |
dc.subject | Croatia | en |
dc.subject | enzyme linked immunosorbent assay | en |
dc.subject | Estonia | en |
dc.subject | fluorescent enzyme immunoassay | en |
dc.subject | France | en |
dc.subject | Germany | en |
dc.subject | Greece | en |
dc.subject | human | en |
dc.subject | Hungary | en |
dc.subject | immunoassay | en |
dc.subject | Italy | en |
dc.subject | laboratory | en |
dc.subject | multicenter study | en |
dc.subject | nephelometry | en |
dc.subject | Netherlands | en |
dc.subject | Norway | en |
dc.subject | quality control | en |
dc.subject | quantitative assay | en |
dc.subject | Spain | en |
dc.subject | Sweden | en |
dc.subject | Switzerland | en |
dc.subject | turbidimetry | en |
dc.subject | United Kingdom | en |
dc.subject | clinical laboratory service | en |
dc.subject | clinical trial | en |
dc.subject | quality control | en |
dc.subject | standard | en |
dc.subject | Autoantibodies | en |
dc.subject | Clinical Laboratory Services | en |
dc.subject | Humans | en |
dc.subject | Laboratories | en |
dc.subject | Quality Control | en |
dc.subject | Reference Standards | en |
dc.subject | De Gruyter Open Ltd | en |
dc.title | Repository of intra- and inter-run variations of quantitative autoantibody assays: A European multicenter study | en |
dc.type | journalArticle | en |
Ficheros en el ítem
Ficheros | Tamaño | Formato | Ver |
---|---|---|---|
No hay ficheros asociados a este ítem. |