dc.description.abstractPurpose: In an attempt to improve field uniformity and CT number stability, the software in a CT scanner was replaced with a new version. The purpose of the present study was to investigate the influence of the software upgrading on density measurements. Material and Methods: The upgrading influence was investigated by means of a torso phantom (European spinal). The phantom simulates a torso which contains 3 vertebrae of different densities. Results: Despite the use of a reference standard under the phantom, the software change resulted in: a) improved short-term precision; b) increased density values at 80 and 100 kVp, decreased at 130 kVp; and c) increased influence of the kVp on the values. Conclusion: The results of the present study suggest that conversion factors have to be used in density values for reference populations as well as in the values of patients undergoing follow-up studies.en
dc.source.urihttp://www.scopus.com/inward/record.url?eid=2-s2.0-0030277284&partnerID=40&md5=974564f1cd0b5e92d135ca97a82b192d
dc.subjectBone mineral densityen
dc.subjectComputed tomographyen
dc.subjectQuantitativeen
dc.subjectRadiation therapyen
dc.subjectSoftwareen
dc.subjectarticleen
dc.subjectcomputer assisted tomographyen
dc.subjectcomputer programen
dc.subjecthumanen
dc.subjectimage qualityen
dc.subjectmethodologyen
dc.subjectradiodensitometryen
dc.subjectradiographyen
dc.subjectspineen
dc.subjectDensitometry, X-Rayen
dc.subjectHumansen
dc.subjectPhantoms, Imagingen
dc.subjectTomography, X-Ray Computeden
dc.titleSoftware upgrading in a CT scanner: Influence on quantitative CTen