Accuracy of a Newly-Introduced Oscillometric Device for the Estimation of Arterial Stiffness Indices in Patients on Peritoneal Dialysis: A Preliminary Validation Study

Thumbnail
Autor
Vaios V., Georgianos P.I., Pikilidou M.I., Eleftheriadis T., Zarogiannis S., Papagianni A., Zebekakis P.E., Liakopoulos V.
Datum
2018
Language
en
Schlagwort
arterial stiffness
blood pressure
blood pressure monitoring
human
oscillometry
peritoneal dialysis
pulse wave
validation study
Blood Pressure
Blood Pressure Monitoring, Ambulatory
Humans
Oscillometry
Peritoneal Dialysis
Pulse Wave Analysis
Vascular Stiffness
NLM (Medline)
Zur Langanzeige
Zusammenfassung
The aim of the present study was to compare the aortic systolic blood pressure (aSBP), heart-rate-adjusted augmentation index (AIx75), and pulse wave velocity (PWV) obtained using the Mobil-O-Graph (IEM, Stolberg, Germany) and SphygmoCor (AtCor, Sydney, Australia) devices in patients receiving peritoneal dialysis (PD).After a 10-minute rest in the supine position, the Mobil-O-Graph and SphygmoCor devices were applied in randomized order in 27 consecutive PD patients. The agreement between the measurements produced by the Mobil-O-Graph and SphygmoCor devices was explored using Bland-Altman analysis.The Mobil-O-Graph-derived aSBP, AIx75, and PWV did not differ from the same measurements obtained with SphygmoCor (aSBP: 120.5 ± 18.2 mmHg vs. 124.4 ± 19.0 mmHg, p = 0.438; AIx75: 27.0% ± 12.4% vs. 24.5% ± 10.6%, p = 0.428; PWV: 9.5 ± 2.1 m/s vs. 10.1 ± 3.1 m/s, p = 0.397). The slight difference in the estimation of aSBP is possibly explained by the difference in brachial SBP used for the calibration of the devices (131.0 ± 20.6 mmHg vs. 134.5 ± 19.7 mmHg, p = 0.525). Mobil-O-Graph-derived measurements correlated strongly with paired measurements obtained with the SphygmoCor device. Bland-Altman plots showed no evidence of asymmetry and a wide range of agreement between the two devices.Our study shows acceptable agreement between Mobil-O-Graph and SphygmoCor in the estimation of arterial stiffness indices in PD patients. Accordingly, the Mobil-O-Graph device accurately performs aortic ambulatory blood pressure monitoring in this population.
URI
http://hdl.handle.net/11615/80330
Collections
  • Δημοσιεύσεις σε περιοδικά, συνέδρια, κεφάλαια βιβλίων κλπ. [19743]
htmlmap 

 

Stöbern

Gesamter BestandBereiche & SammlungenErscheinungsdatumAutorenTitelnSchlagwortenDiese SammlungErscheinungsdatumAutorenTitelnSchlagworten

Mein Benutzerkonto

EinloggenRegistrieren
Help Contact
DepositionAboutHelpKontakt
Choose LanguageGesamter Bestand
EnglishΕλληνικά
htmlmap